If you believe that your paper was mistakenly rejected by other leading journals and you do not agree with final decision, the editors of Reports of Practical Oncology and Radiotherapy offer new fast track review. You may submit your manuscript to Reports of Practical Oncology and Radiotherapy together with all prior peer-reviews obtained from the other journal and your rebuttal letter. We guarantee review based decision within 72 hours from the time we will receive your manuscript.

Fast track submission process: Please submit the manuscript with all reviews and rebuttal letter by email to Dr. Michal Masternak (michal.masternak@ucf.edu) for fast review processing. To assure immediate attention the email title must to include: RPOR-fast track- Last Name First Name (of corresponding author).

Volume 24, Number 6, 2019

Set-up error validation with EPID images: Measurements vs Egs_cbct simulation

D. van Eeden, F.H.J. O’Reilly, F.C.P. du Plessis

Summary:

Aim In this study, the egs_cbct code’s ability to replicate an electronic portal imaging device (EPID) is explored. Background We have investigated head and neck (H&N) setup verification on an Elekta Precise linear accelerator. It is equipped with an electronic portal imaging device (EPID) that can capture a set of projection images over different gantry angles. Methods and materials Cone-beam computed tomography (CBCT) images were reconstructed from projection images of two different setup scenarios. Projections of an Anthropomorphic Rando head phantom were also simulated by using the egs_cbct Monte Carlo code for comparison with the measured projections. Afterwards, CBCT images were reconstructed from this data. Image quality was evaluated against a metric defined as the image acquisition interval (IAI). It determines the number of projection images to be used for CBCT image reconstruction. Results From this results it was established that phantom shifts could be determined within 2 mm and rotations within one degree accuracy using only 20 projection images (IAI = 10 degrees). Similar results were obtained with the simulated data. Conclusion In this study it is demonstrated that a head and neck setup can be verified using substantially fewer projection images. Bony landmarks and air cavities could still be observed in the reconstructed Rando head phantom. The egs_cbct code can be used as a tool to investigate setup errors without tedious measurements with an EPID system.

Signature: Rep Pract Oncol Radiother, 2019; 24(6) : 614-623


« back

 
INDEXED IN:

Indexed in: EMBASE®, the Excerpta Medica database, the Elsevier BIOBASE (Current Awareness in Biological Sciences) and in the Index Copernicus.

http://www.sciencedirect.com/science/journal/15071367/19/2